Key Features
Usability
AdultLife™ Pro Ventilator is a manually operated device that is intuitive and user friendly. This ventilator is O2 Compatible and can be used with an external blender or a direct source of oxygen.
Design
Manufactured in the USA by a FDA-certified medical manufacturing company and designed to treat adults patients suffering from respiratory problems due to COVID-19, this ventilator features an internal HEPA filter, internal pressure release valves set at 40 cm H2O, humidification on patient circuit, and a replaceable filter on the patient circuit. The AdultLife Pro Ventilator is compatible with 22mm corrugated tubing and accessories (filter, humidifier, etc).
Safety
Features FDA required alarms including, high leakage, hypoventilation and disconnect.
Specifications
Performance
- Modes: CPAP, Ventilator
- Peak Inspiratory Pressure (PIP): 0-40 cm H2O
- Positive End Expiratory Pressure (PEEP): 0-20 cm H2O
- Ventilator Rate: 0-35 breaths/minute
- Inspiratory Time: 0-2.5 seconds
- Expiratory Time: 0-5 seconds
- Breath Rater Per Minute: 0-35 breaths/min
- Tidal Volume: 20-1000cc
- Air Flow Range: 0-100 cm H2O (wall air)
- O2 Concentration: 21%-100%
Safety
- Disconnect/Power Off Alarm
- Max PIP Alarm
- PEEP Alarm
- Occlusion Alarm
- Hypoventilation
Power
- 110-240V compatible (US Type B plug)
Standard Accessories
- Test lung (1x)
- Patient circuit (1x)
- 1/4 inch tubing for external pressure sensor (1x)
- Spare hardware (replacement tubing/clamps for simple repairs) (3x)
Optional Accessories & Add-ons
- Additional patient circuit
Product Documents, Manuals & Brochures

Brochure

Product Guide
Regulatory Certificates
Not available at this time
Videos
FAQ
How does the AdultLife Pro Ventilator measure breaths?
Breaths are measured and limited by pressure measurements set by the user. Because the tidal volume delivered at each specific pressure will be subject to patient’s lung compliance, elastance, and airway resistance, adequacy of tidal volume should be evaluated by patient response, including breath sounds, chest rise, SpO2, arterial blood gases, capnogram (etCO2), etc.
Does the AdultLife Pro synchronize to spontaneous breathing?
The AdultLife™ Pro Ventilator allows patients to take unsupported, spontaneous breaths on top of the machine-delivered breaths, but time is the only mechanism by which machine-supported breaths are triggered. The ventilator’s breath cycle will not automatically synchronize with the patient’s spontaneous breathing.
Is the patient circuit reusable?
No. It is recommended that the circuit be replaced after each use.
Studies & Trials
Not available at this time
Package Contents
- 1x Main Unit
- 1x Spare Hardware
- 1x User Manual
- 1x Mechanical tidal-volume gauge
Warranty
Testimonials
Not available at this time